How to Create a Known Defect with a Calibrated Pin or Tungsten Filament
A control sample is a package containing a deliberately created and identified defect used to verify that a bubble-leak setup can reveal the targeted indication.
This tutorial compares two practical approaches: puncturing the package with a calibrated pin and creating a controlled channel in a heat seal with tungsten filament.
What is a package leak-test control sample?
It is a representative package with a known artificial defect. When tested under defined conditions, the expected bubble indication helps confirm that the apparatus, parameters and operator technique can reveal that defect.
A control sample is a system-suitability and parameter-development tool. It does not independently calibrate the instrument or replace complete method validation.
Separate a detectable defect from test noise.
Bubble tests can be affected by trapped air, porous materials, insufficient pressure, limited headspace, obstructed gas paths and operator interpretation.
A known defect provides a reference response that helps the laboratory develop parameters and check the test setup before evaluating unknown samples.

Calibrated pin or tungsten filament?

Calibrated Pin
FAST + ACCESSIBLEPuncture an appropriate package area with a known pin diameter. DVACI's demonstrated tool includes nominal 150 µm and 250 µm pin options.
- Quick preparation
- Minimal equipment
- Easy defect marking
- Applicable to many films

Tungsten Filament
CONTROLLED CHANNELPlace a selected filament across the heat-seal area before sealing, then remove it to leave a channel. A nominal 50 µm filament is commonly demonstrated.
- Small nominal diameter
- Seal-channel simulation
- Useful for heat seals
- Controlled placement
Diameter is not the whole leak: A pin puncture or removed filament does not automatically create a perfectly characterized flow rate. Material recovery, folds, adhesive, product and channel geometry can partially obstruct the gas path.
Create and identify the puncture consistently.
Select the location
Choose an area without product and where bubble release can be clearly observed.
Hold perpendicular
Position the pin perpendicular to the package surface to reduce an angled or elongated puncture.
Puncture carefully
Use controlled force and avoid tearing or enlarging the defect beyond the intended preparation.
Mark and document
Circle the location and record pin, package, date and preparation details.



Create a channel through a heat-sealed area.
Position the filament
Place it across the intended seal path and prevent movement during handling.
Seal the package
Apply the defined thermal sealing conditions with the filament crossing the seal.
Allow stabilization
Let the seal cool or stabilize according to the preparation procedure.
Remove carefully
Withdraw the filament without tearing or distorting the surrounding seal.
Choose the defect that represents the validation question.
You need a fast, accessible puncture in a clearly observable package area and the nominal pin size is suitable for the sensitivity objective.
You need a small channel through a heat-seal area and can control filament placement, sealing and removal consistently.
Document more than the nominal defect diameter.
A reliable control program defines how the sample is created, stored, checked and replaced.
Label the sample
Record package type, lot, defect method, nominal size, position and preparation date.
Confirm repeatability
Verify that the sample produces the expected indication over repeated tests under controlled parameters.
Manage condition
Define storage, inspection, permitted reuse and replacement criteria because defects can change over time.
Porous-package note: For Tyvek or medical paper, choose the defect location carefully and control material breathing according to the current method. Avoid placing a pin where blocking treatments, folds or package contents can obstruct the leak path.
Use the control to develop pressure and observation time.
For ASTM F2096, increase internal pressure gradually until the known defect produces a clear, repeatable bubble indication. Then establish a justified operating pressure and an observation time that reliably reveals that response without unacceptable package damage.
For vacuum immersion methods such as ASTM D3078, the control likewise helps confirm whether the selected differential and observation conditions can reveal the intended defect.

Control Sample FAQ
Does a control sample calibrate the leak tester?
Not by itself. It verifies that the complete setup can reveal the known defect under defined conditions. Instrument calibration and method validation are separate controlled activities.
Is a 150 µm pin hole equivalent to a 150 µm leak channel?
Not necessarily. Film recovery, puncture shape, material thickness and obstruction can change the effective gas path.
Which method is more precise?
Tungsten filament can create a smaller nominal seal channel, but performance depends on sealing and removal. The better method is the one that can be prepared, documented and verified consistently for the package.
Can one control sample be used for every package?
No. The control should represent the relevant package construction, defect location, gas path and sensitivity objective.
Can a control sample be reused?
Only under a defined lifecycle procedure. Repeated pressure, handling, moisture or material relaxation may change its response.